RUSKA 7250, 7250i,
7250xi & 7250LP
Pressure Controller/Calibrator
Users Manual
Page view 0
1 2 3 4 5 6 ... 121 122

Summary of Contents

Page 1 - 7250xi & 7250LP

PN 3952213 November 2010 © 2010 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice. All prod

Page 2

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual vi

Page 3 - 7250 – Change Language

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 6-12 3. Select Zero [F1]. Do NOT press the Calibrate button. When initiating the zeroing command

Page 4

Maintenance Sensor Photocell Zeroing 6 6-13 3. Enter the Zeroing screen by selecting MENU | CALIBRATE. 4. Select the Primary sensor by pressing

Page 5 - Table of Contents

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 6-14 gkn047.bmp Figure 6-8. Correct Adjustment of Photocell Centering Bar 9. Tighten the Allen s

Page 6

Maintenance Optimizing Control 6 6-15 Optimizing Control The performance of the controller may be optimized for certain environments or performan

Page 7 - Contents (continued)

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 6-16 Note Prior to performing any Auto-Tune functions in an effort to improve the controller perfo

Page 8

Maintenance System Software Update Procedure 6 6-17 should indicate a sensor temperature of 50 oC and a duty cycle between 10 to 90%. If operatin

Page 9

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 6-18 RUSKA 7250 Controller Software Upgrade In addition to the main code, the RUSKA 7250 includes

Page 10 - Users Manual

Maintenance Cleaning 6 6-19 Table 6-3. RUSKA 7250 Additional Parts Part Number Description 3878948 7250xi isolation manifold valve body 3915651

Page 11 - List of Tables

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 6-20

Page 12

7-1 Chapter 7 Preparation for Storage & Shipping Introduction W Caution The procedures given in this Chapter must be strictly adhered to in order

Page 13 - List of Figures

vii List of Tables Table Title Page 1-1. Symbols...

Page 14

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 7-2 Fluke has found that corrugated cardboard boxes provide the best packaging exterior. The box

Page 15 - General Information

Preparation for Storage & Shipping Packing Instructions 7 7-3 8. Label the carton with the following labels: THIS SIDE UP, HANDLE WITH CARE,

Page 16 - Symbols Used in this Manual

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 7-4 Shipping Instructions Fluke recommends using air freight for transportation. Surface transport

Page 17 - Features

A-1 Appendix A Summary of Specifications Uncertainty Analysis To perform an Uncertainty Analysis on a measurement device, you must be able to identify

Page 18

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual A-2 Long Term Stability Long term stability defines how the instrument drifts with time. This spec

Page 19 - General Information

Summary of Specifications Uncertainty Analysis A A-3 Combining the Uncertainty Components Combining the Uncertainty Components into an Expanded

Page 20

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual A-4 The following two tables are uncertainty analysis examples for a RUSKA 7250xi based on a three

Page 21 - Theory of Operation

Summary of Specifications Uncertainty Analysis A A-5 The following two tables are uncertainty analysis examples for a single range RUSKA 7250 bas

Page 22 - Electronics Module

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual A-6 Specifications General Specifications Pressure Range: Minimum; 0 to 5 (0 to 700 mbar) Maxim

Page 23 - Theory of Operation

Summary of Specifications Specifications A A-7 Performance Table A-5. Performance Specifications: RUSKA 7250 Pressure Range 5–1000 psi (0.34–68.

Page 24 - Pneumatics Module

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual viii

Page 25

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual A-8 MODEL RUSKA 7250i RUSKA 7250xi MODE GAUGE GAUGE Precision1 From 25% to 100% FS: 0.005% RDG

Page 26

B-1 Appendix B Summary of Error Messages Summary of Error Messages Negative error numbers are from the Standard Commands for Programmable Instruments

Page 27 - Control Strategy

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual B-2 Summary of Error Messages, cont. -315 Configuration Data Lost. The configuration data has bee

Page 28 - A = Actual

ix List of Figures Figure Title Page 2-1. Calibrator Block Diagram ...

Page 29 - Transducer Module

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual x 4-25. Menu | Test | Sweep - Menu...

Page 30

1-1 Chapter 1 General Information Introduction This manual contains operation and routine and preventive maintenance instructions for the RUSKA 7250,

Page 31 - Software

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 1-2 Keep Away from Live Circuits Operating personnel must at all times observe safety regulations.

Page 32

General Information General Information 1 1-3 Table 1-1. Symbols Symbol Description P This equipment meets the requirements of all relevant Europ

Page 33

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 1-4 Universal Power Supply The Calibrators’s universal power supply accepts AC voltages between 10

Page 34

General Information Standard Equipment and Options 1 1-5 Automatic Head Correction The Calibrator automatically corrects for head pressure betwee

Page 35 - Installation

LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use an

Page 36 - Powering Up the Calibrator

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 1-6

Page 37 - Pneumatic Connections

2-1 Chapter 2 Theory of Operation Introduction The Calibrator’s power supply, electronics, pneumatics, and sensor combine to form a complete, stand-al

Page 38

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 2-2 Electronics Module Back-Plane Board The Back-plane Board is used to interconnect all of the pl

Page 39 - Installation

Theory of Operation Electronics Module 2 2-3 Table 2-1. Conversion Factors Symbol Description Conversion Factor InHg inches of mercury (0°C) =

Page 40

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 2-4 Pneumatics Module The Calibrator’s Pneumatics Module varies depending on whether the Calibrato

Page 41 - Local Operation

Theory of Operation Pneumatics Module 2 2-5 BACKPANEL Control Section VACUUM SENSOR Option VREFERENCEPORT TESTPORTZERO ISOLATION APPLYRELEASE EXH

Page 42 - Cancel, Previous

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 2-6 Measure Mode Pneumatics Figures 2-2, 2-3 and 2-4 summarize the Pneumatics of the RUSKA 7250. R

Page 43 - Tutorial

Theory of Operation Control Strategy 2 2-7 Vent Procedure The vent mode is implemented with a fast multi-step procedure. For gauge instruments, t

Page 44 - Figure 4-4. Units Menu

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 2-8 The outer loop is a lower frequency response, analog-digital loop. The outer loop uses a calib

Page 45 - Local Operation

Theory of Operation Transducer Module 2 2-9 0 5 10 15 20 25Time - secondsPressure gkn054.eps Figure 2-5. Pressure Control Transducer Module Quart

Page 46 - Main Menu

7250 – Change Language Hold key for 5 seconds mode enter gkn001.eps

Page 47 - Controlling Pressure

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 2-10 FUSED QUARTZ BOURDON TUBEAXIS OF ROTATIONMIRRORTORSION HINGEFLEXUREPERMANENT MAGNETELECTROMAG

Page 48 - Step/Jog

Theory of Operation Software 2 2-11 where again P is pressure and a, b, and c are coefficients generated during the calibration procedure as disc

Page 49

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 2-12 thermally stable by viewing the Menu | Display screen and verifying that the temperature is a

Page 50

Theory of Operation Software 2 2-13 HsZ Hardware zero correction. (0 to 4095, center is 2048). SsZ Zero corrections scaling factor. Ratio of Hi

Page 51

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 2-14

Page 52

3-1 Chapter 3 Installation Introduction This section of the manual discusses initial installation for the RUSKA 7250. Installing the Calibrator involv

Page 53

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 3-2 Table 3-1. General Specifications and Parameters Parameter Value Model Operating Humidity 5–9

Page 54 - 6180.7500xkPamTorr =

Installation Observing the Calibrator’s Full Scale Rating 3 3-3 Observing the Calibrator’s Full Scale Rating To observe the Calibrator’s full sca

Page 55

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 3-4 Reference Port The reference port is open to atmosphere for gauge measurements or it can be co

Page 56

Installation Pneumatic Connections 3 3-5 SUPPLYSEE USER'S MANUAL TOFAN IS NORMALLY OFF.TURN ON.NOTICE:EXHAUST TEST REFPOWER85-265 VAC50/60

Page 58

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 3-6

Page 59

4-1 Chapter 4 Local Operation Introduction This section of the manual describes operation of the RUSKA 7250 using the front panel. The local interface

Page 60

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 4-2 Cancel, Previous These keys are used to stop, undo, or exit the current operation. The CANCEL

Page 61

Local Operation Tutorial 4 4-3 Figure 4-2 is a menu tree showing the relationship between all the menus in the system. Refer to it for selection

Page 62

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 4-4 gkn015.bmp Figure 4-4. Units Menu 2. Use the rotary knob located to the right of the display

Page 63

Local Operation Tutorial 4 4-5 will appear below the scratch pad. Notice that the upper left corner still shows MEASURE. The Calibrator stays in

Page 64

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 4-6 Simulated Absolute Instruments This RUSKA 7250 is a Gauge and Vacuum (Negative Gauge) instrume

Page 65

Local Operation Controlling Pressure 4 4-7 gkn019.bmp Figure 4-8. Functions in Main Menu Note For Calibrators with pressure ranges greater than

Page 66

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 4-8 Vent The Vent [F3] function is used to rapidly vent the pressure in the system to atmosphere.

Page 67

Local Operation Menu 4 4-9 gkn020.bmp Figure 4-9. Configurations in Menu Menu | Setup Setup is used to configure the system. It includes setting

Page 68

i Table of Contents Chapter Title Page 1 General Information ... 1-1 Int

Page 69 - Remote Operation

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 4-10 gkn021.bmp Figure 4-11. Menu l Setup l Limits - Menu High Limit—User Defined Maximum Pressur

Page 70 - Configuration

Local Operation Menu 4 4-11 the system due to thermal effects. However these thermal effects will stabilize over time. Once the system has therm

Page 71 - Device Messages

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 4-12 Slew Limit The slew limit is used by the system to assure that the slew rate is not exceeded.

Page 72

Local Operation Menu 4 4-13 Ready Tolerance A Ready indication is generated when in control mode and the measured pressure is reading within this

Page 73 - Remote Operation

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 4-14 Key Click The RUSKA 7250 can be configured to emit a “clicking” sound each time a key is pres

Page 74

Local Operation Menu 4 4-15 gkn024.bmp Figure 4-14. Menu | Setup | Units | Edit Name - Menu Use the < [F4] or > [F5] key to highlight the

Page 75

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 4-16 Protocol The Protocol defines which protocol should be used by the remote interface. The opti

Page 76

Local Operation Menu 4 4-17 Menu | Calibrate The Menu | Calibrate command is used to perform the calibrations on all of the sensors utilized by t

Page 77 - Serial Operation

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 4-18 vacuum sensors tend to increase at higher vacuum levels, therefore, the uncertainty in the Ca

Page 78

Local Operation Menu 4 4-19 MaxTime The max time is the maximum time in seconds, including the dwell time, that the Calibrator can spend on one s

Page 79 - Sample Programs

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual ii Standard Equipment and Options ...

Page 80

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 4-20 gkn028.bmp Figure 4-18. Menu | Program | Edit - Menu 5. Using the numeric keypad, enter the

Page 81 - Zero Sequence

Local Operation Menu 4 4-21 gkn029.bmp Figure 4-19. Menu | Program | Edit | Auto - Menu 6. Using the numeric keypad, enter the values for Start

Page 82

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 4-22 5. Use the < [F4] and > [F5] keys to highlight a character. 6. Turn the rotary knob t

Page 83

Local Operation Menu 4 4-23 3. Press Config [F3]. The configuration screen will appear. 4. Press Recall [F2]. The Calibrator will be set to the

Page 84

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 4-24 gkn033.bmp Figure 4-23. Menu l Program l Run - Menu 4. Press Run [F2]. The configuration of

Page 85

Local Operation Menu 4 4-25 Menu | Test—Sweep Test The sweep function can be used to automatically exercise the elastic sensing element of the de

Page 86

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 4-26 Menu | Test—Remote Test The Menu | Test | Remote section of the system is used to perform a n

Page 87 - Example for 7250

Local Operation Menu 4 4-27 gkn038.bmp Figure 4-28. Menu | Test | Controller - Menu Menu | Display The Menu | Display is used to conveniently mo

Page 88

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 4-28

Page 89 - Maintenance

5-1 Chapter 5 Remote Operation Capabilities The RUSKA 7250 can be operated remotely by a computer. Two interfaces are supported: IEEE-488 and RS-232.

Page 90 - WX Warning

Contents (continued) iii Vacuum Sensor—Option... 3-5 4 Local Operat

Page 91 - Calibration

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 5-2 RS-232 The RS-232 interface supports standard serial operation from a computer to a single RUS

Page 92

Remote Operation Device Messages 5 5-3 To configure the remote interface: 1. The remote interface is configured from the Setup | Remote | Menu.

Page 93

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 5-4 ANSI/IEEE 488.2-1987 Command Summary *CLS Clear Status *ESE? Event Status Enable Query *

Page 94

Remote Operation Device Messages 5 5-5 :DATE? Last zero date :TIME? Last zero time :PRESsure2 Case reference sensor :VALue<n> &

Page 95

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 5-6 :ON <number> Set control on band :OVERshoot ON|OFF|1|0 Set Overshoot Mode LIST

Page 96

Remote Operation Device Messages 5 5-7 To zero the unit via the remote interface use the following sequence: CAL:ZERO:INIT Enter Zero Mode CAL:Z

Page 97

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 5-8 Standard Event Status Register (ESR), Standard Event Status Enable Register (ESE) Bit 7 Power

Page 98

Remote Operation 6005 Interface Panel Emulation 5 5-9 6005 Interface Panel Emulation The RUSKA 7250 may be configured to emulate the IEEE-488 com

Page 99

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 5-10 R0 Unlock keyboard R1,R2 Lock keyboard S0,S1,S2,S3 Set pre-programmed units

Page 100 - Sensor Photocell Zeroing

Remote Operation Sample Programs 5 5-11 Units Unit Number Unit Name RUSKA display units 1 Pa kPa 2 kPa kPa 3 Mpa kPa 4 mbar

Page 101 - Maintenance

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual iv Remote/Local Operation...

Page 102 - Centering Tool Here

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 5-12 4, /* GPIB Address of 7010 */ NO_SAD,

Page 103 - Optimizing Control

Remote Operation Sample Programs 5 5-13 /*---------------------------------*/ /* Check for GPIB Interface Errors */ /*------------------

Page 104 - Fan Operation

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 5-14 device = ibdev (0, /* First GPIB Board */ 4,

Page 105

Remote Operation Sample Programs 5 5-15 gotoxy (20, 7); if (rstat == 0) cprintf ("Stable"); else if (rstat

Page 106 - Replacement Parts

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 5-16 ibwrt (device, s, strlen (s)); } /*-------------------------------------------------------

Page 107 - Cleaning

Remote Operation Sample Programs 5 5-17 /* PRES:TOL 0.001 Set control tolerance to 0.001 %FS */ /* OUTP:MODE CONTROL Enter con

Page 108

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 5-18 } /*--------------------------------------------------------------------*/ /* serial_int : re

Page 109 - Chapter 7

Remote Operation Sample Programs 5 5-19 ; outportb (portbase, ch); } /*-----------------------------------------------------------------

Page 110

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 5-20

Page 111 - Packing Instructions 7

6-1 Chapter 6 Maintenance Introduction Very little maintenance is required for the RUSKA 7250. This section of the manual discusses suggested maintena

Page 112 - Shipping Instructions

Contents (continued) v Cleaning... 6-1

Page 113 - Summary of Specifications

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 6-2 3. Press Electrical [F1]. The electronics self test will run and display the results. 4. Pre

Page 114

Maintenance Calibration 6 6-3 Vacuum Pumps Periodic checks of the user’s vacuum pumps for oil levels. It is very highly recommended that a power-

Page 115 - Summary of Specifications

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 6-4 Note The uncertainty of the final calibration must include the uncertainty of the pressure sta

Page 116

Maintenance Calibration 6 6-5 Step 1 1. The zero procedure will start when the reading is in-range and stable or when OK [F6] is pressed. Press

Page 117

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 6-6 3. Using your calibration standard, generate the pressure shown in the Apply column. When th

Page 118 - Specifications

Maintenance Calibration 6 6-7 Vacuum (Negative Gauge) Calibrations Vacuum mode is an available option. The following configuration should be used

Page 119 - Performance

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 6-8 3. If desired, change the Calibrator’s units of measure (see Chapter 4) to match those of the

Page 120

Maintenance Calibration 6 6-9 7. You will use the calibration standard to apply the two vacuum points on the vacuum sensor. 8. The Actual va

Page 121 - Summary of Error Messages

RUSKA 7250, 7250i, 7250xi & 7250LP Users Manual 6-10 6. Use the numeric keypad and the ENTER key to enter a new value. To correct a mistake in t

Page 122

Maintenance Calibration 6 6-11 gkn043.bmp Figure 6-4. Zeroing Menu 4. Do not disturb the instrument while zeroing is in process. 5. Wait for t